<?xml version="1.0" standalone="yes"?>
<?xml-stylesheet type="text/xsl" href="..\..\XSLTemplates\ErrorsStyleSheet.xsl"?>
<Error>
<ErrorNumber>17699</ErrorNumber>
<ErrorFamily>Exposure Head</ErrorFamily>
<ErrorMessage>
HEAD: Laser safety loop has just opened
</ErrorMessage>
<AppliesTo>
<UnnumberedList>
<ListItem><Text>Thermal</Text></ListItem>
</UnnumberedList>
</AppliesTo>
<ServiceInformation>
<DescriptionText><text>
One or more of the safety interlock switches has been opened.
</text></DescriptionText>
<DescriptionText><text>
This error code applies to Thermal Head firmware version 1.34 and above.
</text></DescriptionText>
</ServiceInformation>
<OperatorAction>
<DescriptionText><text>
All the safety interlock switches must be closed before normal operations can
be resumed.
</text></DescriptionText>
</OperatorAction>
<Cause>
One or more of the safety interlock switches has been opened.
</Cause>
<Cause>
This error may be the result of a faulty head sensor.
Please see http://techplanet.creo.com/output/heads/thermal/content/docs/TH1_errors_17669_17667.pdf
</Cause>
<SeeAlso>
<UnnumberedList>
</UnnumberedList>
</SeeAlso>
<OtherLinks>
<Link><LinkAddress>..\..\ShowAllErrors.htm</LinkAddress><LinkTitle>Firmware 0 Error Index</LinkTitle></Link>
<Link><LinkAddress>..\..\ShowAllCommands.htm</LinkAddress><LinkTitle>Firmware 0 Command Index</LinkTitle></Link>
<Link><LinkAddress>..\..\ShowAllNVS.htm</LinkAddress><LinkTitle>Firmware 0 NVS Index</LinkTitle></Link>
</OtherLinks>
<Updated>
Updated: Wed Jul 16 19:33:42 2008 PST
</Updated>
<Links>
</Links>
</Error>